It offers outstanding temperature stability. Equipped with a colored touch screen controller for displaying temperature, humidity, and C02 concentration. It is available with a heat decontamination function and optional UV disinfection. In forced convection mode unit and offers a heating function that prevents condensation on the outer doors.
Chamber volume | 150 L |
Shelves | 3 Pcs |
Heating method | Air-jacketed, PID Control |
Temperature range | RT + 5 ~ 50 ℃ |
Temperature resolution | 0.1 ℃ |
Temperature fluctuation | 0.1 ℃ |
Temperature gradient | ≤ ± 0.5 ℃ |
Temperature uniformity | ± 0.3 ℃ |
Ambient temperature | RT + 5 ~ 30 ℃ |
CO2 Range | 0 ~ 20 % V/V |
CO2 Control resolution | ±0.1% (IR sensor) |
CO2 Deviation | ≤ ±0.3%V/V |
CO2 Recovery | (Door open 30 S, recovery to 5 %) ≤ 3 min |
Temperature recovery | (Door open 30 S, recovery to 37 °C ) ≤ 8 min |
Humidity method | Natural vaporization > 90% |
Voltage | AC 220 V / 50 Hz |
Power consumption | 700 W |
Interior dimension (W × D × H) | 480 × 530 × 610 mm |
Exterior dimension (W × D × H) | 670 × 770 × 880 mm |
Packing dimension (W × D × H) | 850 × 890 × 1070 mm |
Gross weight | 130 kg |
Air jacketed heater
Standard sterilization method is 90° moist heat disinfection and UV
Stainless-steel chamber, shelves and semicircular arcs
Microorganism filter at inlet provides 99% filtration of bacteria and dust as well as supplies 7-inch touch screen with controller
72-hour machine operation record query function
Infrared sensors can keep CO2 concentration stable and uniform when doors open frequently
Stainless-steel chamber, shelves and semicircular arcs
Auto-controller of fan speed to prevent sample damage
Doors temperature controller prevents dewfall on the glass doors
Separate temperature-limiting alarm system for safe usage
Alarm function for temperature difference, over CO2 concentration, doors open time, UV working status, etc
Options are available to connect printer and computer for data recording (Optional) via RS-485 connector