Flame Photometer LX404FP is a high-quality and high-performance instrument with a color touch screen digital display, used for the determination of the concentration of certain metal ions mainly K, Na and Ba. There will be no damage to the equipment as the air supply is oil and moisture-free. The environmental temperature is 10°C to 35°C and relative humidity is ≤ 85%. The flame failures can be easily detected by protection functions for more safe and reliable operation.
Elements detected | K, Na, Ba |
Display value | Concentration value |
Data range | 0.000~9.999 |
Channel quantity | 3 |
Aspiration rate | <6 ml/min |
Response time | <8 s |
Stability | 3% drift over 15 S when continuously aspirating |
Reproducibility | 3% coefficient of variation for 7 consecutive samples |
LOD ppm | K: 0.01ppm Na: 0.01ppm Ba: 6ppm |
Range |
K: 0 - 100 ppm Na: 0 - 160 ppm Ba: 0 - 3000 ppm |
Linearity |
K: 0.195 ppm (0.39 - 3.12) ppm Na: 0.69 ppm (1.15 - 9.2) ppm Ba: 9 ppm |
Interface | USB |
Power supply voltage | (220 ± 22) V |
Frequency | (50/60±1)Hz |
Net weight | 18 kg |
Packing dimensions | 780 x 560 x 390 mm |
Gross weight | 26 kg |
Used widely in the measurement of alkali elements concentration in all sample types in mineral extraction, Oil Industry, Paper Industry, Pharmaceuticals, Soil Analysis, Utilities, Food & Beverage, Chemical manufacturing and Fertilizer and Clinical laboratory testing and Pathological study in medical health.
Air compressor x 1 pcs
Pu pipe (Air) Φ6 x 4 (2 meter)
Latex pipe (Liquid waste) Φ6 (1.5 meter)
Capillary Φ 1.8 x 0.8 x 180 x 1 pcs
Rubber pipe (Liquefied gas) Φ6 (2 meter)
Liquefied gas regulator x 1 pcs
Locking chuck (Liquefied gas) x 1 pcs
Glass cover x 1 pcs
Cover part (Chimney cover) x 1 pcs
Water trap x 1 pcs
Nebulizer assembly
Syringe
Capillary injection
Ignition valve
Air regulating valve
Software Built-in printer