Automated ELISA Processor LX1003AEP

Description

Automated ELISA Processor LX1003AEP is a fully automated laboratory system integrating sample dispensing, incubation, shaking, washing, and microplate reading. It features high-precision pipetting, 8 independent photometric channels, and a 400 -700 nm spectral range for accurate analysis. With two temperature-controlled shaking incubators, a 16-channel washer, 10 filter positions, and bidirectional LIS connectivity supporting HL7, it delivers efficient and consistent laboratory testing.

Specifications

Spectral range
400 -750 nm
Absorbance
0.0 - 4.0 OD
Reading channels
8 Independent photometric channels
Pipetting range
10 - 1000 µl
Pipetting precision
(100 µl) CV ≤ 1%, accuracy ≤ 2.5%
(1000 µl) CV ≤ 0.5%, accuracy ≤ 1%
Linear range
0.000 - 3.000 Abs
Repeatability
CV ≤ 1.0%
Stability
≤ ± 0.003 Abs
Incubation
6 Shaking incubators
Incubation temperature
RT to 60°C
Temperature uniformity
0.5°C
Temperature fluctuation
± 0.5°C
Shaking function
Horizontal circular oscillation, 1- 3 levels adjustable, amplitude 2.5 mm
Cover position
1 (8 Covers)
Heating method
Dry heating with cover
Working temperature
10°C to 40°C
Working humidity
30% - 80%
Washer head
16 Channel (32 pin), double row cleaning, hanging installation with leakage recovery
Washing residual
< 1 µl
Injection uniformity
CV < 1.5%
Reading accuracy
≤ ± 0.02 Abs from 0.0 to 1.0 OD
≤ ± 0.003 Abs from 1.0 to 2.0 OD
Optical filters
4 Standard filters (405 nm, 450 nm, 492 nm, 630 nm);
Total 10 filter positions
Sensitivity
≥ 0.01 mg/l
Channel difference
≤ 0.02 Abs
Liquid level detection
Air pressure level detection, clot detection function
Sample racks
6 x 20 Push-pull sample rack with detection
Sample channel
2 Channels
TIP
1000 µl, 3 Tip racks
Reagent racks
4 Reagent racks, reagent tanks with detection
Software
Above windows 7 system
LIS System
Bi-direction, support HL7 protocol
Result
Software-based absorbance and results review
Power supply
AC 220 V ± 10%, 50/60 Hz
External dimension & Net weight
Instrument: 1273 x 730 x 1650 mm; 226 Kg
Base stand: 1273 x 750 x 777 mm; 98 Kg
Packing dimension & Gross weight
Instrument: 1450 x 920 x 1130 mm; 310 Kg
Base stand: 1450 x 920 x 1020 mm; 180 Kg
Accessories: 1070 x 900 x 510 mm; 90 Kg
Features
  • Pressure-based liquid detection with tip and clot detection
  • High-precision pipetting with a minimum volume of 10 µl
  • Two independently controlled injection channels with Y-Z movement
  • Disposable tips to minimize cross-contamination
  • Intelligent gripper with real-time position sensing
  • Adjustable gripper spacing and strength for different microplates
  • Dual-thread control for simultaneous, interference-free operation
  • Eight independent temperature-controlled shaking incubation modules with three-level shaking
  • Single independent microplate washers for efficient washing
  • Built-in high-precision microplate reader
  • Single and dual-wavelength detection with 10 filter positions
  • UV Lamp for ultraviolet illumination
  • LED Light for illumination and visual indication
  • IAP Function for integrated automated processing
  • Wide absorbance range with high sensitivity and excellent repeatability
  • Stable performance for automated ELISA testing
Applications
  • Widely used in hospital laboratories, disease prevention and control centers, blood centers, blood banks, animal and plant inspection and quarantine institutions, scientific research facilities, and university laboratories.
Optional Accessories
  • Printer
  • Scanner
  • HEPA Filter
  • Negative pressure working zone