Benchtop High Temperature Drying Oven LX104DOA

Description

Benchtop High-Temperature Drying Oven LX104DOA is a benchtop design, back heating, horizontal force convection with dynamic self-diagnostic technology. This technology helps prevent any failures or damage to the equipment and makes detection easy. Detailed purpose: glassware desiccation, laboratory sample, food, chemical material’s heat denaturalization, heat induration, heat inteneration, moisture eliminated, biomedical engineering instrument sterilization, electronic component’s Torrefaction and aging.

Specifications

Chamber volume
250 L
Control
PID microprocessor
Display
LED display
Convection method
Horizontal forced convection (Fan)
Temperature control range
Room temperature +10℃~400℃
Temperature accuracy
±1℃ at 100℃
Temperature fluctuation
±1.5℃
Temperature uniformity
±2.5℃ at 100℃
Timer
1min-99hr
Shelf size
590 x 570
Shelf quantity (std. /max.)
2/5 pcs
Power
3200W
Power supply
220V / 50Hz
Internal dimensions
600 x 600 x 700 mm
External dimensions
940 x 820 x 925 mm
Net weight
125 kg
Packing dimensions
1090 x 960 x 1070 mm
Gross weight
152 kg
Features
  • (Internal material) 304 stainless steel working chamber and (external material) powder-coated steel housing
  • LED display, soft touch, PID microprocessor
  • Having stainless steel shelf with adjustable shelf
  • Temperature controller with high-speed and high-performance CPU chip
  • High-sensitivity and high precision PT100 as temperature sensor
  • Preset programmed control with a timing range up to 99 hours for power-on, power-off and working
  • Multiple alarms for sensor failure, high/low-temperature deviation, over temperature
  • Can store the parameters memory
  • Temperature display calibration and dynamic self-diagnostic, control technology
Applications
  • Used in industry, mining, laboratories and scientific research for desiccation, Torre faction, wax-melting and sterilization.
Optional Accessories
  • Independent temperature control and alarming system to provide dual protections and stop working in case of over-temperature
  • Inert gas port
  • Inert gas flow meter