16 L Table top Sterilizer LX232SS is an easy-to-use, front-loading, with LCD display Class B autoclave. Built-in rapid steam generator ensures quick sterilization cycles. Displays two selectable temperatures (either 121°C or 134°C).Equipped with drying function, post-sterilization and safety features d. A mechanical and electrical safety feature prevents the door from being opened until the chamber has reached 0 PSI. Having 3 trays for maximizing space in the chamber.
| Sterilizing chamber volume | 16 L |
|---|---|
| Chamber size | φ250 x 350 mm |
| Number of trays | 3 |
| Sterilization temperatures | 121℃/134℃ |
| Product dimension | 560 x 450 x 410 mm |
| Net weight | 48 kg |
| Package dimension | 650 x 520 x 520 mm |
| Gross weight | 55 kg |
| Programme | Temperature | Sterilization time | Total time | Pressure |
| Solid | 134°C | 4 mins | 20 ~ 45 mins | 210 KPa |
| 121°C | 20 mins | 35 ~ 60 mins | 110 KPa | |
| Liquid | 134°C | 10 mins | 35 ~ 60 mins | 210 KPa |
| 121°C | 30 mins | 35 ~ 60 mins | 110 KPa | |
| Wrapped | 134°C | 4 mins | 45 ~ 65 mins | 210 KPa |
| 121°C | 20 mins | 50 ~ 75 mins | 110 KPa | |
| Textile | 134°C | 4 mins | 45 ~ 65 mins | 210 KPa |
| 121°C | 20 mins | 50 ~ 75 mins | 110 KPa | |
| Prion | 134°C | 18 mins | 45 ~ 75 mins | 210 KPa |
| B and D test | 134°C | 3.5 mins | 22 ~ 35 mins | 210 KPa |
| Helix test | 134°C | 3.5 mins | 22 ~ 35 mins | 110 KPa |