80 L Table Top Sterilizer LX276SS
Description
80 L Table Top Sterilizer LX276SS is a Class B pulsating vacuum type autoclave and designed with an automatic microprocessor system with LCD color touch screen display. It has a fully automatic route for water inlet, pre-heating, pulse vacuum, heating, sterilizing, draining, exhausting and vacuum drying. It has various types of Sterilization cycles such as Preset cycles (Liquid, Instrument, Textile, and Rubber) + User-defined cycles + test cycles (Vacuum test, B-D test). The vacuum rate can reach up to -0.096Mpa, it’s suitable for various unwrapped or wrapped instruments, packages, unpackaged, solid, hollow, multi-apertures and pipe systems, dressing, liquid, and glassware. It is mainly used in CSSD, pharmaceutical factories, labs, operation rooms, research institutions and industrial sterilization, etc.
Specifications
| Chamber volume |
80 L |
|---|---|
| Chamber size (Dia. × D) |
Ø386 x 700 mm |
| Control |
Microcomputer |
| Display |
4.3 inch LCD color touch screen display |
| Design pressure |
-0.1/0.3 Mpa |
| Working pressure |
0~0.23Mpa |
| Working temperature range |
105 - 138℃, adjustable |
| Sterilization time |
0 - 99 min |
| Drying time |
0 - 99 min |
| Vacuum rate |
-0.096 Mpa |
| Temperature accuracy |
0.1°C |
| Heat average |
≤1°C |
| Voltage |
220V, 50Hz |
| Power |
5.7 KW |
| Outer dimension |
980 x 800 x 820 mm |
| Net weight |
120 kg |
| Packing dimension |
1020 x 690 x 620 mm |
| Gross weight |
150 kg |
Features
- Fully automatic microprocessor control system with 4.3-inch touch screen LCD display
- It has vacuum drying system
- Safe door lock system ensuring safety/protection during sterilization
- Chamber, door water tank, steam generator, and pipeline are made of high-quality SS304 stainless steel and the outer shell is made up of stainless steel sprayed
- Having unique pulse vacuum drying procedure
- Built-in high-quality vacuum pump to ensure the removal of the air to get optimum saturated steam
- Came with advanced features like water level cut-off sensor, water level indicator, fault warning system, 0.22µm HEPA filter, automatic program control, automatic drain, preset pre-vacuum and post-vacuum
- The door is manually operated and activated by a dedicated button
- Steam-water inner circulation system
- Designed with various types of safety functions such as over-temperature protection, over-pressure protection, automatic release safety valve, overheating protection device, vacuum pump overload and emergency stop button
- It has sound-light alarm when the sterilization process is finished and cuts off the heating power source automatically
- Designed with separate boiler with high-efficiency steam generator to provide very precise control of heating temperature
Applications
- It is commonly used in medical health care institution to make the sterilization for wrapped or unwrapped instruments, fabric, utensils, culture medium, unsealed liquid, etc.
Standard Accessories
- Built-in printer