Fully Automatic Kjeldahl Analyzer LX102FKA is an advanced system for nitrogen and protein analysis based on the Kjeldahl method. It integrates automated distillation and titration with a microcomputer control system for precise operation. The analyzer features a precision dosing pump, real-time condensation temperature monitoring, and a large LCD touchscreen for accurate and efficient laboratory performance.
| Display |
LCD Touch screen |
|---|---|
| Measurement range |
0.1 - 240 mg N |
| Distillation time |
5 - 10 Min / sample |
| Recovery rate |
≥ 99.5 % |
| Sample |
Solid ≤ 5 g, liquid ≤ 20 ml / sample |
| Condensate water consumption |
1.5 L / Min |
| Titration precision |
2.0 µl |
| Repeatability RSD |
≤ 0.5 % |
| Data storage capacity |
1000 Groups |
| Power |
2000 W |
| Power supply |
AC 220 V ± 10 %, 50 / 60 Hz |
| External dimension |
455 x 391 x 730 mm |
| Net weight |
38 Kg |
| Packing dimension and gross weight |
590 x 530 x 910 mm; 75 Kg 610 x 600 x 360 mm; 17 Kg |