Horizontal Autoclave LX241HA uses saturated steam as the sterilizing medium, operating on a pulsating vacuum principle to ensure effective and uniform sterilization. It features a robust rectangular chamber design with a single horizontal sliding door and a double-channel door sealing system, integrated with a temperature-controlled door safety interlock for enhanced operational safety. It is equipped with a PLC and a full-color touch screen interface, enabling fully automatic program control and precise monitoring.
| Capacity |
1500 L |
|---|---|
| Inner chamber dimension (W x D x H) |
680 x 1800 x 1180 mm |
| Design temperature |
150°C |
| Design pressure |
1.0 MPa |
| Chamber design pressure |
- 0.1/0.3 MPa |
| Jacket design pressure |
0.3 MPa |
| Working temperature range |
139°C |
| Working pressure range |
0.25 MPa |
| Temperature accuracy |
0.1°C |
| Pressure accuracy |
0.1 KPa |
| Temperature control accuracy |
± 1°C |
| Temperature uniformity |
± 1°C |
| Chamber relief valve |
0.28 MPa |
| Jacket relief valve |
0.28 MPa |
| Maximum vacuum |
- 96 KPa |
| Negative pressure pulsating vacuum |
- 80 to 0 KPa |
| Across the pressure pulsation |
-80 to 80 KPa |
| Positive pressure pulse |
0 - 9 Times |
| Steam consumption |
750 Kg/C |
| Water source consumption |
80 Kg/C |
| Control |
PLC |
| Power supply |
380 V, 50 Hz, 4 kW (90 kW) |
| Outer dimension |
2357 x 2110 x 2113 mm |
| Net weight |
2450 Kg |