Horizontal Autoclave LX272HA

Description

Horizontal Autoclave LX272HA is a double-door, rectangular, horizontal and triple-layer design fully automatic controlled autoclave that disinfects and sterilizes objects with saturated steam. It has a vertical sliding door with an effective sterilization temperature of 150 °C. Having safety features (like auto door, automatic shutoff preventing current overload, overpressure and temperature protective device, etc.), automatic drainage control system and steam generator user can also set up new cycles using the user-defined cycle to modify the parameters of all preset cycles as needed.

Specifications

Capacity
100 L
Inner chamber dimension (L × W × H)
450 × 450 × 550 mm
Design temperature
150 ℃
Design pressure
-0.1/0.3Mpa
Working temperature range
105 ℃ - 139 ℃
Working pressure range
0 - 0.25Mpa
Temperature control accuracy
≤±0.2 ℃
Temperature display accuracy
0.1 ℃
Sterilization time range
0-999min
Drying time range
0-999min
Vacuum rate
-97Kpa
Temperature equilibrium
≤±0.5 ℃
Safety valve open pressure
0.28Mpa
Steam Pressure and the consumption
0.3 - 0.6 MPa and 8 kg/cycle
Water Source Pressure and the consumption
0.15 - 0.3 MPa and 60 kg/cycle
Compressed air Pressure
0.4 - 0.7 MPa
Control
PLC + HMI + Printer
Interface
USB, Ethernet
Power supply
12 kW
Outer dimension (L × W × H)
1100 x 1700 x 1000 mm
Net weight
450 Kg
Packing dimension
1280 x 1900 x 1180 mm
Gross weight
630 Kg
Features
  • Front loading, 7-inch color touch screen display to monitor working status and parameters 
  • Motorized or sliding door with pneumatic sealing to prevent loss of sterile steam 
  • Built-in PLC + HMI+ printer 
  • Built-in standard steam generator 
  • Having 6 preset sterilization cycles, 4 User-defined cycles, and Testing cycles (B&D test and vacuum leakage test) and the user can create more new cycles if needed 
  • Contains an aluminum silicate insulating layer for the best insulation technique 
  • Equipped with an independent temperature monitoring system, the controller system has four-level password access, and a pneumatic angle pipeline system with removable filter plugs to prevent clogging 
  • Perfect fault information prompt and fault self-diagnosis function 
  • Door safety interlock feature: The door cannot be opened till the pressure inside the chamber doesn’t fall to 0 
  • Automatic power cut off with alarm with low water levels indication
Applications
  • Used in medical, research, microbiological, pharmaceutical, food industry and clinical departments for sterilization and also to sterilize the waste before discarding.
Programmed Specifications
ProgrammeTemperatureSterilization timeDrying time
Unwrapped instruments134°C6 mins8 mins
Wrapped instruments134°C7 mins15 mins
Textile program134°C7 mins15 mins
Implant program134°C10 mins15 mins
Liquid program121°C20 minsNo
Rubber121°C20 mins15 mins
Hollow134°C6 mins10 mins
Prions134°C18 mins15 mins
PCD134°C3.5 mins15 mins
Bowie-Dick test134°C3.5 mins5 mins
Standard Accessories
  • Loading shelves x 2 pcs
Optional Accessories
  • Remote monitoring system 
  • Noise reduction configuration 
  • Data logger 
  • 10-inch touch screen 
  • ERP system