Horizontal Autoclave LX276HA

Description

Horizontal laboratory autoclave LX276HA is a double-door, rectangular, horizontal and triple-layer design fully automatic controlled autoclave that disinfects and sterilizes objects with saturated steam. It has a motorized or sliding door with an effective sterilization temperature of 150 °C. Having safety features (like auto door, automatic shutoff preventing current overload, overpressure and temperature protective device, etc.), automatic drainage control system and steam generator user can also set up new cycles using the user-defined cycle to modify the parameters of all preset cycles as needed.

Specifications

Capacity
300 L
Inner chamber dimension (L × W × H)
550 × 550 × 1000 mm
Design temperature
150 ℃
Design pressure
-0.1/0.3Mpa
Working temperature range
105 ℃ - 139 ℃
Working pressure range
0 - 0.25Mpa
Temperature control accuracy
≤±0.2 ℃
Temperature display accuracy
0.1 ℃
Sterilization time range
0-999min
Drying time range
0-999min
Vacuum rate
-97Kpa
Temperature equilibrium
≤±0.5 ℃
Safety valve open pressure
0.28Mpa
Steam Pressure and the consumption
0.3 - 0.6 MPa and 18 kg/cycle
Water Source Pressure and the consumption
0.15 - 0.3 MPa and 140 kg/cycle
Compressed air Pressure
0.4 - 0.7 MPa
Control
PLC + HMI + Printer
Interface
USB, Ethernet
Power supply
22 kW
Outer dimension (L × W × H)
1290 × 1850 × 1350 mm
Net weight
820 kg
Packing dimension
1490 × 2050 × 1550 mm
Gross weight
1020 Kg
Features
  • Front loading, 7-inch color touch screen display to monitor working status and parameters
  • Motorized or sliding door with pneumatic sealing to prevent loss of sterile steam
  • Built-in PLC + HMI+ printer
  • Built-in standard steam generator
  • Having 6 preset sterilization cycles, 4 User-defined cycles, and Testing cycles (B&D test and vacuum leakage test) and the user can create more new cycles if needed
  • Contains an aluminum silicate insulating layer for the best insulation technique
  • Equipped with an independent temperature monitoring system, the controller system has four-level password access, and a pneumatic angle pipeline system with removable filter plugs to prevent clogging
  • Perfect fault information prompt and fault self-diagnosis function
  • Door safety interlock feature: The door cannot be opened till the pressure inside the chamber doesn’t fall to 0
  • Automatic power cut off with alarm with low water levels indication
Applications
  • Used in medical, research, microbiological, pharmaceutical, food industry and clinical departments for sterilization and also to sterilize the waste before discarding.
Programmed Specifications
Programme Temperature Sterilization time Drying time
Unwrapped instruments 134°C 6 mins 8 mins
Wrapped instruments 134°C 7 mins 15 mins
Textile program 134°C 7 mins 15 mins
Implant program 134°C 10 mins 15 mins
Liquid program 121°C 20 mins No
Rubber 121°C 20 mins 15 mins
Hollow 134°C 6 mins 10 mins
Prions 134°C 18 mins 15 mins
PCD 134°C 3.5 mins 15 mins
Bowie-Dick test 134°C 3.5 mins 5 mins
Standard Accessories
  • Loading cart x 1 pcs
  • Transfer trolley x 2 pcs
Optional Accessories
  • Remote monitoring system
  • Noise reduction configuration
  • Data logger
  • 10-inch touch screen
  • ERP system