Microplate Reader LX770MR

Description

Microplate Reader LX770MR is used in ultraviolet-visible spectrophotometer measurement of 96 well microplates in 10 seconds and 384 well microplates in 30 seconds, it contains linear plate shaking motion. Equipped with a xenon lamp, long life and stable illumination. It offers free wavelength selection from 200-1000nm, so is an ideal tool for virtually any photometric research application, such as DNA, RNA, protein analysis and more it offers exceptional usability through its intuitive, powerful PC Reader It-II Software. It has a special light absorption detection board that can be used for calibration of the equipment also used in data analysis software, microplate analysis and workstations.

Specifications

Wavelength range
200 nm to 1000 nm with 1 nm steps
Wavelength accuracy
±2 nm
Optical system
Monochromator
Bandwidth
˂ 2.5 nm
Repetition
±0.2 nm
Measurement range
0 to 4.0 OD
Detector
Two silicon photo detectors, one for measurement, another for reference
Linearity at 450 nm
R2≥0.999, (0.0-3.0 Abs)
Absorbance accuracy at 450nm
±(1.0 % + 0.003 Abs), (0~2.0 Abs), ± 2.0 % [2.0 ~ 2.5 Abs)
Absorbance repeatability at 450nm
CV
Test speed
10 seconds with 96-well plate (fast mode), 30 seconds with 96-well plate (accurate mode) using end point method
Plate shaking
Linear, three options for speed
Temperature range
RT+4°C~45°C
Temperature accuracy
±0.5°C @37°C
Temperature uniformity
±0.1°C @37°C
Ambient emperature
+5℃ to 45℃
Software
Integrated software or PC control software
Light source
Xenon flash lamp/number of flashes>10°
Display
10 inch touch screen
Analysis software
Reader It- II
User interface
Touch screen, Android system, 10 inch touch screen, External keyboard mouse
Storage
16 GB memory, more than 20,000 test records can be stored
Ports
3 USB ports for PC, Printer and network port
Automated systems
Temporarily unable compatible with automated systems
Power supply
DC 24 V 6.5 A
Dimension
300 x 500 x 260 mm
Net weight
15 kg
Packing dimension
670 x 460 x 450 mm
Gross weight
19 kg
Features
  • Build with temperature control incubation system
  • Equipped with a grating monochromator
  • Instrument parameter setting and instrument self-test function are highly automated
  • There are 3 shaking speeds:- low speed (5Hz, amplitude: 15mm), intermediate speed (11 Hz, amplitude: 3mm) and high speed (20Hz, amplitude: 1mm)
  • Heating up time from 25℃ to 37℃ is less than 45 min
  • Powerful PC software (Reader It -II), to offer analysis review, validation of data
  • It is easy access to data via a USB interface and Wi-Fi
  • It is suitable for cell multiplication, gene report, enzymology, immune, cytotoxicity, cell apoptosis and GPCR detection
  • Having the stepper motor and incubator shut down to save energy
  • CCD real-time spectral scanning mode and real-time output of UV-visible spectrum
  • It measures 96 and 384 well microplates
  • Oscillating mixing function
Applications
  • Used in especially DNA/RNA analysis, cell proliferation and cytotoxicity test, bacterial concentration analysis, pharmaceutical plant analysis and used in endpoint method ELISA / EIA, with a variety of fitting curves for analysis.