Nucleic Acid Extraction System LX103NAE
Description
Nucleic Acid Extraction System LX103NAE automates nucleic acid extraction and purification using magnetic bead-based separation. It processes 1 - 24 samples per run, supports cfDNA extraction with a maximum volume of 8 mL, and is ideal for research labs, diagnostic facilities, and the biotechnology industry.
Specifications
| Throughput |
1 - 24 Samples |
|---|---|
| Processing volume |
50 µl - 8 ml |
| Temperature accuracy |
1.5°C |
| Heating temperature |
37°C - 125°C |
| Plate/Strip capacity |
Up to 24 reagent strips or 6 plates |
| Processing time |
20 - 40 Minutes per run |
| Reagent type |
Magnetic bead-based reagent kits (Open system) |
| Mixing speed (rod sleeves) |
0 - 3000 rpm |
| Mixing method |
Rotational mixing (rod sleeves) + Linear shaking (plate) |
| Magnetic bead recovery rate |
98 % |
| Inter-well purification variance |
CV < 3 % |
| Data storage capacity |
> 500 Programs |
| Noise |
< 60 dB |
| Display |
10 Inch color touch screen |
| UV sterilization |
Built-in HEPA filtration system |
| Power supply |
AC 100 - 240 V, 50/60 Hz |
| Extension ports |
2 USB Ports |
| External dimension (W x D x H) |
349 x 440 x 388 mm |
| Net weight |
32 Kg |
| Packing dimension (W x D x H) |
630 x 500 x 580 mm |
| Gross weight |
50 Kg |
Features
- Large volume processing up to 8 mL with elution as low as 50 µl
- Ideal for cfDNA and pathogenic DNA extractions, including wastewater
- Advanced vortex (3000 RPM) and linear plate mixing for thorough sample lysis
- Built-in UV lamp and filtration system to prevent cross-contamination
- Rapid heating from RT to 100°C in 1 minute (37°C - 125°C range)
- Compact, space-saving design for high-throughput workflows
Applications
- Used for automated extraction of DNA and RNA from blood, plasma, tissue, swabs, and cultured cells, supporting PCR, RT-PCR, qPCR, and NGS workflows. Ideal for molecular diagnostics, clinical laboratories, pathogen detection, genetic testing, and genomic research.
Standard Accessories
- Reagent strips (Single sample)
- Strip plates (4 Samples)
- Magnetic rod sleeves