Oxygen, Nitrogen And Hydrogen Analyzer LX200ONH
Description
Oxygen, Nitrogen and Hydrogen Analyzer LX200ONH is an advanced benchtop instrument designed for simultaneous determination of oxygen, nitrogen and hydrogen in metals, alloys and inorganic materials using the inert gas fusion principle. It offers high sensitivity, precise detection and fast analysis with a typical testing time of approximately 3 minutes. Equipped with advanced detection systems and software, it is ideal for quality control, metallurgical testing, steel plants, foundries and research laboratories.
Specifications
| Elements analysis |
Oxygen (O), Nitrogen (N), Hydrogen (H) |
|---|---|
| Oxygen measuring range |
Low range: 0.00001 - 0.5% High range: 0.5 - 50% |
| Nitrogen measuring range |
Low range: 0.00001 - 0.5% High range: 0.5 - 30% |
| Hydrogen measuring range |
Low range: 0.00001 - 0.15% High range: 0.00001 - 0.15% (IR / TCD) |
| Range extension |
Up to 100% by adjusting sample weight |
| Oxygen detection |
Non-dispersive infrared (NDIR) |
| Nitrogen detection |
High-precision thermal conductivity detector (TCD) |
| Hydrogen detection |
NDIR or TCD (Optional) |
| Oxygen precision |
Low range: 0.1 ppm or 1% High range: 2.0 ppm or 1% |
| Nitrogen precision |
Low range: 0.1 ppm or 1% High range: 2.0 ppm or 1% |
| Hydrogen precision |
0.2 ppm or 2% |
| Analysis channels |
16 High-range and 16 low-range channels per element |
| Sensitivity |
0.01 ppm |
| Cooling system |
Dedicated recirculating water cooling |
| Operating temperature |
20 - 25°C |
| Relative humidity |
≤ 70% RH |
| Analysis time |
Approx. 3 minutes per sample |
| Sample type & weight |
Solid samples, chips and suitable inorganic materials with 0.01- 1 g |
| Furnace control |
Pulse electrode furnace; maximum furnace current 1500 A; power 10 kVA; temperature above 3500°C |
| Heating programs |
Constant-power, ramp and multi-step |
| Calibration |
Single-point, multi-point and blank correction |
| Connectivity |
Computer, printer, electronic balance and supported database integration |
| Software |
Windows-based interface |
| Power supply |
220 V AC ±10%, 50 Hz |
| Packing dimension & Gross weight |
Main unit: 710 x 810 x 960 mm; 215 Kg Water chiller: 650 x 560 x 840 mm; 100 Kg Accessories: 710 x 810 x 970 mm; 130 Kg |
Features
- Inert gas fusion principle for accurate elemental analysis
- Fast analysis time of approximately 3 minutes per sample
- Advanced NDIR detection for oxygen and hydrogen
- High-precision TCD detection for nitrogen
- High-temperature pulse electrode furnace above 3500 °C
- Programmable heating with constant-power, ramp and multi-step modes
- Automatic leak detection and real-time gas flow monitoring
- Temperature-controlled detectors for improved stability and precision
- Open-entry furnace design for easy sample loading and cleaning
- Automatic high/low measuring range switching
- Flexible calibration with blank correction and multi-point calibration
- Windows-based software with live curves and automatic data storage
- PDF and Excel reporting with statistical analysis
Applications
- Used for simultaneous oxygen, nitrogen and hydrogen analysis in steel, cast iron, non-ferrous alloys, titanium, zirconium and other inorganic materials. It is ideal for quality control, metallurgical testing, foundries, industrial laboratories and materials research.
Standard Accessories
- Electronic balance x 1 Set
- Analysis software x 1 Set
- water chiller x 1 Set
- Voltage stabilizer
- Computer x 1 Set
- Printer x 1 Set