Semi-Auto Coagulation Analyzer LX200ACA
Description
Semi-Auto Coagulation Analyzer LX200ACA is a compact, benchtop unit with optical colorimetry methodology and 2 test channels. Designed with an advanced program self-test and user-friendly interface with LCD display, it provides a quality control analysis function with real-time reaction curve monitoring and ensures test results reliability. With a built-in thermal printer with an RS-232 serial port, it has memory storage for up to 200000 pieces of data with analysis reports in an international standard format.
Specifications
| Testing principle |
Optical colorimetry |
|---|---|
| Testing channel |
2 (can test different projects at the same time) |
| Display |
5 |
| Cuvette incubation positions |
12 |
| Reagent incubation positions |
5 |
| Wavelength |
470 nm |
| Detection time |
Normal sample takes 20 to 40 sec and can be set random from 20 to 90 sec |
| Incubation time |
1-999S |
| Reaction time |
1-99S |
| Sample type |
Plasma |
| Results |
20,000 results storage, automatically saves measurement data when powered off |
| Reagent dosage |
Minimum dosage 20μl |
| Sample volume |
20-40μl |
| Incubation temperature |
37.0±1.0℃ |
| Interface |
RS232 serial interface, support LIS system |
| Printer |
Built-in thermal printer |
| Relative humidity |
≤80% |
| Precision |
PT, APTT, TT, FIB |
| Ambient temperature |
10°~30°C |
| Atmospheric pressure |
86KPa~106KPa |
| Power supply |
70VA, AC100~240V, 50/60Hz |
| External dimension |
360 x 210 x 120 mm |
| Net weight |
3.5 Kg |
| Packing dimension |
380 x 250 x 320 mm |
| Gross weight |
5 Kg |
Features
- 5- Inch LCD touch screen display
- The pipette automatically triggers the measurement and the measurement time can be tenths of a second
- Having an open system, good compatibility, supports reagents of various specifications
- It can test several parameters like PT, APTT, TT, and FIB in plasma
- Storage data up to 20,000 pieces of data
- Equipped with multiple pre-temperature positions to improve detection efficiency
- Built-in thermal printer with RS-232 serial port for easy printing
- Advanced program self-test with a user-friendly interface
- Optical method detector
- LIS system available
- Less reagent consumption
Applications
- Used to measure a coagulation pathway speed, as well as thrombin and thromboplastin levels, and for detecting blood coagulation factors in medical care, scientific research education institutions, etc.
Reagent Parameters
| Item | Sample volume | Reagent 1 | Reagent 2 | Incubating time | CV% |
| PT | 20μl | 40μl | / | 120 S | ≤5% |
| APTT | 20μl | 40μl | 20μl | 180 S | ≤5% |
| TT | 30μl | 30μl | / | 120 S | ≤8% |
| FIB | 40μl | 20μl | / | 180 S | ≤5% |