Hematology analyzer LX301HMA 3 part differentiation fully automatic unit that can carry out 60 tests per hour. It features a built-in thermal printer and it automatically calculates CV%, SD and saves data. Stores up to 100,000 results including histograms. It is used widely in patient and research. All the function are high efficient and easy and fast to operate.
Throughput | 60 tests/hour | |||||||||||||||
Measuring Principle | Impedance for cell counting Cyanide-free method for HGB | |||||||||||||||
Aperture Diameter | WBC chamber:100μm,RBC/PLT chamber:70μm | |||||||||||||||
Parameter | WBC, Neu#, Lym#, Mid#, Neu%,Lym%, Mid%, RBC, HGB, HCT, MCV, MCH, MCHC,RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, P-LCR | |||||||||||||||
Histograms | 3 histograms (includingWBC, RBC, PLT histogram) | |||||||||||||||
Sampler Volume | Normal:10μl; Predileted:20μl | |||||||||||||||
Calibration | Manual & auto calibration | |||||||||||||||
Counting Modes | Venous, capilary, prediluted | |||||||||||||||
Printer | Built-in thermal printer,USB to external printer(optional) | |||||||||||||||
Input and output | RS232 port, USB port, VGA port. LIS system avaliable | |||||||||||||||
Option | Barcode scanner, LIS | |||||||||||||||
Storage | 100,000 results with histogram | |||||||||||||||
Carry-over rate |
| |||||||||||||||
Background |
| |||||||||||||||
Precision |
| |||||||||||||||
Power supply | AC 220±10%, 60/50Hz, 110±10%, 60Hz | |||||||||||||||
Package Size (W x D x H) | 560 x 470 x 600 mm | |||||||||||||||
Gross weight | 24 kgs |
Measuring Principle is Impedance for cell counting Cyanide-free method for HGB
It has manual & auto calibration system
Equipped with touch screen display, internal printer and optional external printer
Easy to operate, Friendly and Intellective Software
It has three counting modes
Designed with 3-part differentiation for detecting the parameters of RBC, WBC, HGB and differential counting
RS232 port, USB port, VGA port. LIS system available
Storage capacity up to 100,000 data with histogram