Whatsapp
Hematology Analyzer LX301HMA

Hematology Analyzer LX301HMA

Hematology analyzer LX301HMA 3-part differentiation fully automatic unit that can carry out 60 tests per hour. It features a built-in thermal printer and it automatically calculates CV%, and SD also saves data. Stores up to 100,000 results including histograms. It is used widely in patients and research. All the functions are highly efficient, easy and fast to operate.

Catalog
Throughput60 tests/hour
Measuring principleImpedance for cell counting Cyanide-free method for HGB
ReagentLyse(500mL) and Diluent(20L)
ParameterWBC, Neu#, Lym#, Mid#, Neu%, Lym%, Mid%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, P-LCR, NLR, PLR, PLCC
Histograms3 histograms (including WBC, RBC, PLT histogram)
Sampler volumeVenous mode: 10μl venous blood Capillary mode: 10μl capillary blood Pre-diluted mode: 20μl L capillary blood
CalibrationManual, Auto and fresh blood calibration
Counting modesVenous, capillary and pre-diluted
Quality control3-level QC, L-J graph
Temperature18℃-35℃
PrinterThermal printer, support external printer
Input and outputRS232 port, USB port, VGA port. LIS system avaliable
OptionBarcode scanner, LIS
Storage100,000 results with histogram
Carry-over rate
Parameter CV
WBC ≤ 2%
RBC ≤ 1.9%
HGB ≤ 1.9%
MCV ≤ 0.4%
PLT ≤ 4.0%
Precision
Parameter Measurement range CV
WBC (0-99.9×109)/L (7.0 -15.0×109)/L
RBC (0-9.99×1012)/L (3.5 - 6.0×1012)/L
HGB (0-300.0)g/L (110.0 - 180.0)g/
PLT (0-999×109)/L (100.0 - 5 00.0×109)/L
Power supplyAC 100-240V, 50/60±1Hz
Dimension (W x D x H)410 x 435 x 472 mm
Net weight21 kg
Package dimension (W x D x H)560 x 470 x 600 mm
Gross weight24 kg

Measuring Principle is Impedance for cell counting Cyanide-free method for HGB

It has a manual, auto and fresh blood calibration system

Equipped with 8- inch touch screen display, printer and optional external printer

Easy to operate, Friendly and Intellective software

It has three counting modes

Designed with 3-part differentiation for detecting the parameters of RBC, WBC, HGB and differential counting

One press for sample analysis with a throughput of 60 sec per test

High voltage shock and back-flush to remove a blockage, auto as well as manual print is supported

RS232 port, USB port, VGA port. LIS system available

Storage capacity up to 100,000 data with histograms

Hematology analyzers are used to run tests on blood samples. They are used in the medical field to do white blood cell counts, complete blood counts, reticulocyte analysis, and coagulation tests.

Related Products

//