Automated NGS Library Preparation System LX102NGS is a fully automated workstation designed for high-precision DNA and RNA sequencing library preparation. It automates the entire workflow from fragmentation and end repair to adaptor ligation and PCR amplification for up to 96 samples. The system also features integrated HEPA filtration, UV sterilization, and intuitive drag-and-drop software to eliminate cross-contamination and ensure precise results. Built-in thermal cycling ensures highly accurate fluid handling across various reagents.
| Throughput (96 Samples) | DNA Library preparation: 96 Samples RNA Library preparation: 32 Samples Target capture: 16 Samples Single cell sequencing: 24 Samples |
|---|---|
| Pipetting range | 1 - 200 µL |
| Pipetting precision | 1 µL: CV ≤ 8% 2 µL: CV ≤ 5% 5 µL: CV ≤ 4% 20 µL: CV ≤ 1% 100 µL: CV ≤ 1% 200 µL: CV ≤ 1% |
| Pipetting accuracy | 1 µL ± 10% 2 µL ± 8% 5 µL ± 4% 20 µL ± 2% 100 µL ± 1% 200 µL ± 1% |
| Minimum detection volume | 20 µL |
| Tiles number | 27 |
| Channel | Flexible 24 - channel pipetting system |
| Compatibility | Deep well plates and PCR plates |
| Working temperature | RT 20 ± 5°C |
| Relative humidity | 88% |
| Temperature control block | Including 2 temperature control block |
| Temperature control range | 4 - 105°C |
| Temperature control accuracy | ≤ 1°C |
| Temperature control uniformity | ≤ 1°C at 55°C |
| Thermal cycling block temperature range | Thermo lid: 30°C to 120°C Block: 4°C to 99°C |
| Thermal cycling block uniformity | ≤ 0.7°C |
| Thermal cycling block Accuracy | ≤ ± 0.3°C |
| Fluorescence detection throughput | 24 |
| Fluorescence detection range | 0.4 ng - 100 ng |
| Consumable stacking mode | Stack plate lifting block |
| Magnetic plate | 96-Well annual magnetic plate |
| Functional module | UV Sterilization lamp HEPA Filtration |
| Information management | External scanner for sample/reagent information LIS Connectivity Cloud database connectivity |
| Interface | USB Port |
| Power | 1200 W |
| Power supply | 100/240 V, 50/60 Hz |
| External dimension | 1420 x 790 x 800 mm |
High-precision pipetting for up to 96 samples
Intelligent error detection and resource availability alerts
Adjustable liquid handling for accurate aspiration and dispensing
Integrated PCR, fluorometer, heating mixing block and gripper
Customizable modules for different library preparation workflows
Automated sample-to-library preparation
Drag-and-drop workflow programming
Easy-to-use and editable GUI
Guided operation for precise workflow execution
Negative-pressure HEPA filtration and UV sterilization
PCR Sealing options to minimize condensation and cross-contamination
It's used for automated preparation of DNA and RNA sequencing libraries for NGS applications. Ideal for High-throughput sample processing for genomic, transcriptomic, and molecular research workflows.