Automated ELISA Processor LX1005AEP
Description
Automated ELISA Processor LX1005AEP is a fully automated laboratory system integrating sample dispensing, incubation, shaking, washing, and microplate reading. It features high-precision pipetting, 8 independent photometric channels, and a 400-700 nm spectral range for accurate analysis. With two temperature-controlled shaking incubators, a 16-channel washer, 10 filter positions, and bidirectional LIS connectivity supporting HL7, it delivers efficient and consistent laboratory testing.
Specifications
| Spectral range |
400 -750 nm |
|---|---|
| Absorbance |
0.0 - 4.0 OD |
| Reading channels |
8 Independent photometric channels |
| Pipetting range |
5 - 1000 µl |
| Pipetting precision |
(100 µl) CV ≤ 1.0%, accuracy ≤ 2.5% (1000 µl) CV ≤ 0.5%, accuracy ≤ 1.0% |
| Linear range |
0.000 - 3.000 Abs |
| Repeatability |
CV < 1.5% |
| Stability |
≤ ± 0.003 Abs |
| Incubation |
16 Shaking incubators |
| Incubation temperature |
RT to 60°C |
| Temperature uniformity |
± 0.5°C |
| Temperature fluctuation |
± 0.5°C |
| Shaking function |
Horizontal circular oscillation, 1-3 levels adjustable, amplitude 2.5 mm |
| Cover position |
4 (16 Covers) |
| Heating method |
Dry heating with cover |
| Working temperature |
10°C to 30°C |
| Working humidity |
40% - 70% |
| Washer head |
16 Channels (32 pin), double row cleaning, hanging installation, with leakage recovery |
| Washing residual |
< 1 µl |
| Injection uniformity |
CV < 1.5% |
| Reading accuracy |
≤ ± 0.02 Abs from 0.0 to 1.0 OD ≤ ± 0.03 Abs from 1.0 to 2.0 OD |
| Optical filters |
4 Standard filters (405 nm, 450 nm, 492 nm, 630 nm) Total 10 filter positions |
| Sensitivity |
≥ 0.01 mg/l |
| Channel difference |
≤ 0.02 Abs |
| Liquid level detection |
Air pressure level detection, needle blocking intelligent detection |
| Sample racks |
12 x 20 Push pull sample rack with detection |
| Sample channel |
8 Channels |
| TIP |
5 TIP racks |
| Reagent racks |
6 Reagent positions, 40 quality control positions + 12 reagent bottles + 25 reagent tanks (reagent tank with code identification) |
| Software |
Above windows 7 system |
| LIS System |
Bi-direction, support HL7 protocol |
| Result |
Software-based absorbance and results review |
| Power supply |
AC 100 V/240 V, 50/60 Hz |
| External dimension & Net weight |
Main unit: 1925 x 824 x 1652 mm; 320 Kg Base stand: 1925 x 792 x 777 mm; 153 Kg |
| Packing dimension & Gross weight |
Main unit: 2100 x 950 x 1160 mm; 430 Kg Base stand: 2100 x 950 x 1010 mm; 255 Kg Accessories: 1560 x 930 x 530 mm; 130 kg |
Features
- Pressure-based liquid detection with tip and clot detection
- High-precision pipetting with a minimum volume of 5 µl
- 8 Independently controlled injection channels with Y-Z movement
- Disposable tips to minimize cross-contamination
- Intelligent gripper with real-time position sensing
- Adjustable gripper spacing and strength for different microplates
- Dual-thread control for simultaneous, interference-free operation
- 16 Independent temperature-controlled shaking incubation modules with three-level shaking
- Three independent microplate washers for efficient washing
- Built-in high-precision microplate reader
- Single and dual-wavelength detection with 10 filter positions
- UV Lamp for ultraviolet illumination
- LED Light for illumination and visual indication
- IAP Function for integrated automated processing
- Wide absorbance range with high sensitivity and excellent repeatability
- Stable performance for automated ELISA testing
Applications
- Widely used in hospital laboratories, disease prevention and control centers, blood centers, blood banks, animal and plant inspection and quarantine institutions, scientific research facilities, and university laboratories.
Optional Accessories
- Printer
- Scanner
- HEPA Filter
- Negative pressure working zone