Horizontal Laboratory Autoclave LX169HA
Description
Horizontal laboratory autoclave LX169HA is a double-door, rectangular, horizontal and triple-layer design fully automatic controlled autoclave that disinfects and sterilizes objects with saturated steam. It has a vertical sliding door with an effective sterilization temperature of 150 °C. Having safety features (like auto door, automatic shutoff preventing current overload, overpressure and temperature protective device, etc.), automatic drainage control system and steam generator user can also set up new cycles using the user-defined cycle to modify the parameters of all preset cycles as needed.
Specifications
| Capacity |
600 L |
|---|---|
| Sterilization temperature |
150℃ |
| Sterilization pressure |
-0.1/0.3Mpa |
| Working temperature range |
105℃ - 139℃ |
| Working pressure range |
0 - 0.25Mpa |
| Temperature control accuracy |
≤±0.2℃ |
| Temperature display accuracy |
0.1℃ |
| Sterilization time range |
0-999min |
| Drying time range |
0-999min |
| Vacuum rate |
-97Kpa |
| Temperature equilibrium |
≤±0.5℃ |
| Safety valve open pressure |
0.28Mpa |
| Steam Pressure and the consumption |
0.3 - 0.6 MPa and 35 kg/cycle |
| Water source pressure and the consumption |
0.15 - 0.3 MPa and 220 kg/cycle |
| Compressed air pressure |
0.4 - 0.7 MPa |
| Control |
PLC + HMI + Printer |
| Interface |
USB, Ethernet |
| Inner chamber dimension (L × W × H) |
652 × 690 × 1350 mm |
| Outer dimension (L × W × H) |
1400 × 1950 × 1700 mm |
| Power supply |
39 kW |
| Net weight |
1150 kg |
| Packing dimension |
1600 × 2150 × 1900 mm |
| Gross weight |
1350 Kg |
Features
- Front loading, 7-inch colour touch screen display to monitor working status and parameters
- Motorized or sliding door with pneumatic sealing to prevent loss of sterile steam
- Built-in PLC + HMI+ printer
- Built-in standard steam generator
- Having 6 present sterilization cycles, 4 User-defined cycles, and Testing cycles (B&D test and vacuum leakage test) and the user can create more new cycles if needed
- Contains an aluminium silicate insulating layer for the best insulation technique
- Equipped with an independent temperature monitoring system, the controller system has four-level password access, and a pneumatic angle pipeline system with removable filter plugs to prevent clogging
- Perfect fault information prompt and fault self-diagnosis function
- Door safety interlock feature: The door cannot be opened till the pressure inside the chamber doesn’t fall to 0
- Automatic power cut off with alarm with low water levels indication
Applications
- Used in medical, research, microbiological, pharmaceutical, food industry and clinical departments for sterilization and also to sterilize the waste before discarding.
Programmed Specifications
| Programme | Temperature | Sterilization time | Drying time |
| Unwrapped instruments | 134°C | 6 mins | 8 mins |
| Wrapped instruments | 134°C | 7 mins | 15 mins |
| Textile program | 134°C | 7 mins | 15 mins |
| Implant program | 134°C | 10 mins | 15 mins |
| Liquid program | 121°C | 20 mins | No |
| Rubber | 121°C | 20 mins | 15 mins |
| Hollow | 134°C | 6 mins | 10 mins |
| Prions | 134°C | 18 mins | 15 mins |
| PCD | 134°C | 3.5 mins | 15 mins |
| Bowie-Dick test | 134°C | 3.5 mins | 5 mins |
Standard Accessories
- Loading cart x 1pcs
- Transfer trolley x 2pcs
Optional Accessories
- Remote monitoring system
- Noise reduction configuration
- Data logger
- 10-inch touch screen
- ERP system