TOC Analyzer LX101TOC is applied for the determination of TOC (Total Organic Carbon Content) present in liquid or gas samples by the “high-temperature catalytic combustion oxidation Method”. It is equipped with oxygen (O2) gas and a non-dispersive infrared (NDIR) detector for the determination of carbon in the range of 0 to 50000 mg/l. It offers personalized standard curve management, real-time sample measurement, consumables management, easy-to-retrieve measurement records and 680°C catalytic oxidation technology with a platinum catalyst.
| Measurement Range |
0-1000 mg/l, which can be extended to 50,000mg/L by manual dilution |
|---|---|
| Detection Limit |
TC (standard catalyst): 50 μg/l TC (platinum aluminum catalyst: 100 μg/l IC: 20 μg/l |
| Detection method |
High-temperature catalytic combustion oxidation |
| Detector |
NDIR |
| Parameters |
TC, IC, TIC, TOC, NPOC |
| Sample type |
Liquid Sample |
| Gas Requirement |
Oxygen ≥ 99.995% |
| Repeatability |
0.03 |
| Measuring time |
TC: approx. 4 min IC: approx. 3 min |
| Injection volume |
TC: 100-500µL IC: 100-2000µL |
| Operation Mode |
PC Control |
| Power |
1W, AC220V, 50Hz or AC110V, 60Hz |
| Weight |
45 kg |
| Packing dimension |
780 × 710 × 610 mm |
| Gross weight |
60 kg |