45 L Table Top Sterilizer LX274SS is a Class B pulsating vacuum type autoclave and designed with an automatic microprocessor system with LCD color touch screen display. It has a fully automatic route for water inlet, pre-heating, pulse vacuum, heating, sterilizing, draining, exhausting and vacuum drying. It has various types of Sterilization cycles such as Preset cycles (Liquid, Instrument, Textile, and Rubber) + User-defined cycles + test cycles (Vacuum test, B-D test). The vacuum rate can reach up to -0.096Mpa, it’s suitable for various unwrapped or wrapped instruments, packages, unpackaged, solid, hollow, multi-apertures and pipe systems, dressing, liquid, and glassware. It is mainly used in CSSD, pharmaceutical factories, labs, operation rooms, research institutions and industrial sterilization, etc.
Chamber volume | 45 L |
Chamber size (Dia. × D) | Ø316 x 590 mm |
Control | Microcomputer |
Display | 4.3 inch LCD color touch screen display |
Design pressure | -0.1/0.3 Mpa |
Working pressure | 0~0.23Mpa |
Working temperature range | 105 - 138℃, adjustable |
Sterilization time | 0 - 99 min |
Drying time | 0 - 99 min |
Vacuum rate | -0.096 Mpa |
Temperature accuracy | 0.1°C |
Heat average | ≤1°C |
Voltage | 220V, 50Hz |
Power | 3.6 KW |
Outer dimension | 850 x 600 x 520 mm |
Net weight | 90 kg |
Packing dimension | 860 x 690 x 620 mm |
Gross weight | 110 kg |
Fully automatic microprocessor control system with 4.3-inch touch screen LCD display
It has vacuum drying system
Safe door lock system ensuring safety/protection during sterilization
Chamber, door water tank, steam generator, and pipeline are made of high-quality SS304 stainless steel and the outer shell is made up of stainless steel sprayed
Having unique pulse vacuum drying procedure
Built-in high-quality vacuum pump to ensure the removal of the air to get optimum saturated steam
Came with advanced features like water level cut-off sensor, water level indicator, fault warning system, 0.22µm HEPA filter, automatic program control, automatic drain, preset pre-vacuum and post-vacuum
The door is manually operated and activated by a dedicated button
Steam-water inner circulation system
Designed with various types of safety functions such as over-temperature protection, over-pressure protection, automatic release safety valve, overheating protection device, vacuum pump overload and emergency stop button
It has sound-light alarm when the sterilization process is finished and cuts off the heating power source automatically
Designed with separate boiler with high-efficiency steam generator to provide very precise control of heating temperature