Dissolved Gas Analyzer LX100DGA is designed with an advanced three-detector system, enabling complete analysis of the total dissolved gas content in insulating oil using a single sample injection. It accurately detects key gases including hydrogen, methane, ethylene, ethane, acetylene, carbon monoxide, and carbon dioxide, supporting effective condition monitoring of insulating oils.
| Detector | FID, TCD | ||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Minimum detection concentration (μl/L) |
|
||||||||||||||
| Sampler | Packed column sampling, Capillary sampling, Six-port valve gas sampling | ||||||||||||||
| Display | 8 Inch color LCD touch screen | ||||||||||||||
| Temperature Control Range | Above room temperature, 4°C - 450°C | ||||||||||||||
| Temperature Increment | 1°C | ||||||||||||||
| Temperature Accuracy | ± 0.1°C | ||||||||||||||
| Temperature Control Area | 8 Channels | ||||||||||||||
| Programmed Temperature Rising Rate | 0.1 - 60°C/Min | ||||||||||||||
| Sampling Start | Manual | ||||||||||||||
| FID Detection Limit | ≤ 5 x 10ˉ¹² g/s (Cetane/Iso octane ) | ||||||||||||||
| FID Baseline noise | ≤ 0.07 PA | ||||||||||||||
| FID Baseline drift | ≤ 0.2 PA/30 Min | ||||||||||||||
| FID Linear range | ≥ 10⁶ | ||||||||||||||
| TCD Sensitivity | S ≥ 10000 mV x ml/mg (Benzene/Toluene) | ||||||||||||||
| TCD baseline noise | ≤ 20 μV | ||||||||||||||
| TCD baseline drift | ≤ 30 μV/30 Min | ||||||||||||||
| TCD linear range | ≥ 10⁴ | ||||||||||||||
| Interface | Ethernet, IEEE802.3 | ||||||||||||||
| Power supply | 220 V, 50 Hz | ||||||||||||||
| External dimension | 690 x 610 x 490 mm | ||||||||||||||
| Net weight | 55 Kg | ||||||||||||||
| Packing dimension | 790 x 690 x 740 mm 720 x 600 x 540 mm 570 x 440 x 460 mm |
||||||||||||||
| Gross weight | 170 Kg |
Maximum 3 detectors FID 2, TCD 1
Injection volume 1 mL; minimum detection concentration specified per gas
Equipped with 8-inch color LCD touch screen
Integrated 8 temperature control channels
Manual or automatic sampling start
Supports 8 external channels with 2 auxiliary control outputs
Features an Ethernet port (IEEE 802.3) for connectivity
Used for analyzing the total content of dissolved gas components in Petroleum and oil Industries.
Air generator x 1 Pc
Hydrogen generator x 1 Pc
Oscillator x 1 Pc
Auto-sampler
Labdex Multitest Analyzer performs multiple biochemical tests simultaneously with precision and efficiency, making it an essential instrument for clinical and research laboratories aiming to optimize workflow.
Labdex Semi-Automatic Chemistry Analyzers deliver flexible, precise biochemical testing for modern laboratories. Designed for control and efficiency, they support routine and specialized assays with semi-automatic operation. Ideal for clinical and research labs, they enhance workflow, accuracy, and effective diagnostic performance.
Labdex Grain Moisture Meter ensures accurate moisture measurement in grains to prevent spoilage and maintain quality. A must-have tool for agricultural labs, grain storage, and processing industries.
Labdex Automatic Biochemistry Analyzer delivers fast, accurate testing for liver, kidney, lipid, and glucose profiles. With full automation, LIS connectivity, and minimal maintenance, it's ideal for high-throughput clinical labs.
Labdex Blood Gas and Electrolyte Analyzer delivers precise and rapid analysis of blood gases and electrolytes essential for critical care diagnostics. Designed for accuracy and ease of use, this analyzer supports efficient patient management in hospitals and laboratories.