Horizontal laboratory autoclave LX187HA is a single-door, horizontal pulsating vacuum and fully automatic controlled autoclave that disinfects and sterilizes objects with saturated steam. It has a motorized or sliding door with an effective sterilization temperature of 150 °C. Having safety features (like auto door, automatic shutoff preventing current overload, over pressure and temperature protective device, etc.), automatic drainage control system, steam generator and the user can set up new cycles using the user-defined cycle to modify the parameters of all preset cycles as needed.
Capacity | 150 L |
Sterilization temperature | 150℃ |
Sterilization pressure | -0.1/0.3Mpa |
Working temperature range | 105℃-134℃ |
Working pressure range | 0-0.25Mpa |
Temperature control accuracy | ≤±0.2℃ |
Temperature display accuracy | 0.1℃ |
Sterilization time range | 0-999min |
Drying time range | 0-999min |
Vacuum rate | -97Kpa |
Temperature equilibrium | ≤±0.5℃ |
Safety valve open pressure | 0.28Mpa |
Steam Pressure and the consumption | 0.3 - 0.6 MPa and 12 kg/cycle |
Water source pressure and the consumption | 0.15 - 0.3 MPa and 80 kg/cycle |
Compressed air pressure | 0.4 - 0.7 MPa |
Inner chamber dimension (L × W × H) | 450 × 450 × 750 mm |
Outer dimension (L × W × H) | 1200 × 1100 × 1850 mm |
Power supply | 2 kW Control system + 12 kW Steam generator |
Net weight | 550 kg |
Packing dimension | 1400 × 1280 × 2050 mm |
Front loading, 7-inch color touch screen display to monitor working status and parameters
Motorized or sliding door with pneumatic sealing to prevent loss of sterile steam
Built-in PLC + HMI + printer
Built-in standard steam generator
Contains an aluminum silicate insulating layer for the best insulation technique
Equipped with an independent temperature monitoring system, the controller system has four-level password access, and a pneumatic angle pipeline system with removable filter plugs to prevent clogging
Perfect fault information prompt and fault self-diagnosis function
Door safety interlock feature: The door cannot be opened till the pressure inside the chamber doesn’t falls to 0
Automatic power cut off with alarm with low water levels indication
Related Products