Horizontal laboratory autoclave LX270HA is a double-door, rectangular, horizontal and triple-layer design fully automatic controlled autoclave that disinfects and sterilizes objects with saturated steam. It has a vertical sliding door with an effective sterilization temperature of 150 °C. Having safety features (like auto door, automatic shutoff preventing current overload, overpressure and temperature protective device, etc.), automatic drainage control system and steam generator user can also set up new cycles new cycles using the user-defined cycle to modify the parameters of all preset cycles as needed.
Capacity | 120 L |
Inner chamber dimension (L × W × H) | 450 × 450 × 650 mm |
Design temperature | 150 ℃ |
Design pressure | -0.1/0.3Mpa |
Working temperature range | 105 ℃ - 139 ℃ |
Working pressure range | 0 - 0.25Mpa |
Temperature control accuracy | ≤±0.2 ℃ |
Temperature display accuracy | 0.1 ℃ |
Sterilization time range | 0-999min |
Drying time range | 0-999min |
Vacuum rate | -97Kpa |
Temperature equilibrium | ≤±0.5 ℃ |
Safety valve open pressure | 0.28Mpa |
Steam Pressure and the consumption | 0.3 - 0.6 MPa and 10 kg/cycle |
Water Source Pressure and the consumption | 0.15 - 0.3 MPa and 70 kg/cycle |
Compressed air Pressure | 0.4 - 0.7 MPa |
Control | PLC + HMI + Printer |
Interface | USB, Ethernet |
Power supply | 14 kW |
Outer dimension (L × W × H) | 1190 × 1700 × 1000 mm |
Net weight | 520 kg |
Packing dimension | 1390 × 1900 × 1180 mm |
Gross weight | 700 Kg |
Front loading, 7-inch color touch screen display to monitor working status and parameters
Motorized or sliding door with pneumatic sealing to prevent loss of sterile steam
Built-in PLC + HMI+ printer
Built-in standard steam generator
Having 6 preset sterilization cycles, 4 User-defined cycles, and Testing cycles (B&D test and vacuum leakage test) and the user can create more new cycles if needed
Contains an aluminum silicate insulating layer for the best insulation technique
Equipped with an independent temperature monitoring system, the controller system has four-level password access, and a pneumatic angle pipeline system with removable filter plugs to prevent clogging
Perfect fault information prompt and fault self-diagnosis function
Door safety interlock feature: The door cannot be opened till the pressure inside the chamber doesn’t fall to 0
Automatic power cut off with alarm with low water levels indication
Programme | Temperature | Sterilization time | Drying time |
Unwrapped instruments | 134°C | 6 mins | 8 mins |
Wrapped instruments | 134°C | 7 mins | 15 mins |
Textile program | 134°C | 7 mins | 15 mins |
Implant program | 134°C | 10 mins | 15 mins |
Liquid program | 121°C | 20 mins | No |
Rubber | 121°C | 20 mins | 15 mins |
Hollow | 134°C | 6 mins | 10 mins |
Prions | 134°C | 18 mins | 15 mins |
PCD | 134°C | 3.5 mins | 15 mins |
Bowie-Dick test | 134°C | 3.5 mins | 5 mins |