Vertical Autoclave LX141VA is a sterilizing mechanism with top loading sterilizer along with safety protection features like an LCD display system, heat protection, Anti-dry heating, safety interlock door, over temperature, pressure and electric protection, Eight double inner locks, lid closing checking, dry scorch protection, Automatic release safety valve. Electric safety protection device and emergency exhaust switch for a safe and effective sterilizing process. It adopts a working environment of 5-45℃ and relative humidity of 10% - 85%. This mechanism is very effectively useful in hospitals, clinical laboratories, research centers, and science colleges for the sterilization of glass wares, biological materials, cultural media, surgical instruments, rubber material, etc.
Capacity | 54 Liters |
Chamber dimension | φ325 x 725 mm |
Stainless steel basket | (φ300 x 200)mm x 2 |
Temperature accuracy | 0.1℃ |
Working environment | 5-45℃, relative humidity 10% - 85% |
Display | LCD screen |
Chamber material | AISI316L |
Sterilization temperature | 105 - 138℃ |
Range of sterilization time | 1-6000 min |
Dissolving temperature | 40-115℃ |
Range of dissolving time | 1-6000 min |
Warming temperature | 40-79℃ |
Range of warming time | 1-9999 min |
Drying | 1-300 min |
Cooling lock open | 40-80℃ for liquid cycle, 40-99℃ for other cycles |
Sterilization cycle number | Can save up to 60 cycles, record 1000 processes, and USB port |
Exhaust | Automatic internal discharging Exhaust speed can be selected by 6 levels |
Opening way | Up and down, one-touch open |
Preset range of timer | Can be preset and auto-start up to 15 days in advance |
Range of pressure gauge | -0.1-0.6Mpa |
Pressure | Max working pressure: 0.5Mpa Design pressure: 0.61Mpa |
Controller | Fast-speed microcomputer controller |
Interface | USB |
Power supply | 2600 W, 220V, 16A, 50/60HZ |
Dimension | 460 x 560 x 1030 mm |
Net weight | 110 kg |
Package dimension | 560 x 660 x 1230 mm |
Gross weight | 125 kg |
It has an Electric double inner lock for -time monitoring of the opening and closing state
Having the lid locking system through the split column with one key unlock
Background security test
The system monitors the temperature changes in the sterilization chamber in real time
It has a large screen LCD display
Sterilization record storage system which can store 1000 sterilization process
Sterilization through six different exhaust speeds can be selected according to the substances
Sterilization data can be converted into curves through software
Equipped with the cooling process at the end of the sterilization chamber
It can be manually controlled with 0.5 Mpa pressure control
It has two steam collectors with fast speed microprocessor controller
Built-in F0 value calculation function
Related Products