Table Top Spectrophotometer LX300TS is a dual optical path design based on a differential spectroscopy engine and 1nm grating spectrum analysis technology which monitors the energy fluctuation and interference of light source during the measurement of sample signal guaranteeing the instrument measuring accuracy and the long-term repeatability index up to dE*ab ≤ 0.01. Equipped with an Inter-instrument agreement and provides two kinds of light sources. It adopts reflectance D/8° and transmittance D/0° geometry to measure both opaque and transparent materials with automatic calibration technology that confirms the long-term uniformity of fluorescence measurements.

Catalog
Specifications
Wavelength range 360 nm to 780 nm
Wavelength pitch 10 nm
Sensor Differential spectrum engine
Grating method Concave grating
Sphere diameter 152 mm
Half bandwidth 1.6 nm
Reflectance range 0 to 200 %
Resolution 0.0001
Light source Pulse xenon lamps and LED
UV measurement Including UV, 400nm cutoff, 420nm cutoff, 460nm cutoff
Operating temperature 5°C to 40°C
Storage temperature 20°C to 45°C
Relative humidity 80% RH (at 35°C) below no condensation
Measurement time SCI/SCE < 2s (Single mode)
SCI+SCE < 4s
Measurement aperture reflectance XLAV Φ25.4mm/Φ30mm
LAVΦ15mm/Φ18mm, MAVΦ8mm/Φ11mm
SAVΦ3mm/Φ6mm
Transmittance: Φ17mm/Φ25mm (Auto aperture size recognition)
Viewing angles 2° and 10° Standard observer
Transmittance sample size No limit on sample width and height
Thickness ≤ 50mm
Repeatability ΔE*ab≤0.01
Spectral reflectance/transmittance≤0.1%
Long-term repeatability XLAV chroma value: standard deviation ΔE*ab within 0.1 (0℃-40℃ arbitrary temperature change)
XLAV chroma value: standard deviation ΔE*ab 0.01 or less (under constant temperature conditions, the white correction plate is measured every hour within 24 hours)
Inter-instrument agreement XLAV ΔE*ab 0.18
Illuminants A, B, C, D50, D55, D65, D75, F1, F2, F3, F4, F5, F6, F7, F8, F9, F10, F11, F12, CWF, U30, U35, DLF, NBF, TL83, TL84, ID50, ID65, LED-B1, LED-B2, LED-B3, LED-B4, LED-B5, LED-BH1, LED-RGB1, LED-V1, LED-V2
Display Spectral data, Spectrogram, chromaticity data, color difference Data, chromaticity map, pass/fail Judgment, Simulation color, Color evaluation, fog, liquid chromaticity, Color bias
Color space CIE LAB, CIE LUV, LCh, Hunter Lab, Yxy, XYZ, Musell, s-RGB, βxy
Chromatic aberration formula ΔE*ab, ΔE*CH, ΔE*uv, ΔE*cmc, ΔE*94, ΔE*00, ΔEab(hunter), 555 color classification
Chromaticity index WI (ASTME31320, ASTME31373, CIE, ISO2470/R457, AATCC, Hunter, Taube, Berger stensby), YI(ASTMD1925, ASTM E313-20, ASTM E313-73), Tint(ASTM E313-20), Isochromatic index milm, color fastness, color changing fastness, ISO brightness, R457, A density, T density, E density, M density, APHA/Hazen/Pt-Co(platinum-cobalt index), Gardner(Gardner Index), Saybolt(Seibert Index), Astm color, fog, total transmittance, covering power, force, intensity
Lighting/light receiving system Reflection: d/8(diffuse lighting, 8° direction reception), transmission:d/ 0 (diffuse lighting, vertical direction reception)
SCI(including specular reflected light)/SCE(not including specular reflected light) measurement at the same time
Compliance with: CIE No.15, GB/T 3978, GB 2893, GB/T 18833, ISO7724/1, DIN5033 Teil7, JIS Z8722 Conditions C, ASTM E1164, ASTM-D1003-07
Interface USB, USB-B, RS-232, Bluetooth
Screen size 7 Inches capacitive touch screen
Storage memory 8 GB U Disk for data storage and transfer
Power 12V/3A
Camera positioning Ultra HD camera (1400dp)
Automatic calibration Can greatly improve the long-term repeatability of the instrument
Fluorescence calibration Can automatically adjust the UV intensity, and ensure that the value of the instrument is highly consistent with that of other imported instruments when measuring materials containing fluorescence
Brightness calibration Through the brightness calibration algorithm, the real color of ultra-dark samples is restored
Other functions PC side software save sample image function
The instrument can measure sideways, upwards and downwards (with accessories)
Automatic temperature and humidity compensation function
Dimensions (L × W × H) 650 x 450 x 450 mm
Weight 11 kg
Features
  • 7-inch capacitive touch screen, Android operating system

  • 37 standard light sources and 40+measurement indicators

  • Using the MEMS process makes the color measurement more accurate on the basis of 1nm spectral resolution

  • Self-developed fluorescence calibration technique

  • Recording visual data while collecting data for obtaining further details on the sample

  • Dual optical paths spectrum analysis technology

  • Compatibility with SCI/SCE measurement system

  • Inter-instrument agreement

  • 4 apertures switching with automatic recognition and allows the measurement of samples having many shapes

  • Four types of UV test modes

  • Support PC end export or printing of data test reports

Applications
  • Used for the color and transmittance measurement in industrial samples like petroleum and oil, textile and yarn, pellet materials and packing material, etc.

Standard Accessories
  • Power adapter

  • USB cable

  • Fixture for transmittance

  • Software U disk

  • Black cavity

  • White and Green board

  • 30mm, 18mm, 11mm and 6mm apertures

  • Support table

  • Colorimetric dish

Optional Accessories
  • Heating transmission fixture (including control circuit)

  • Vertical bracket

  • Pneumatic jacking rod (including control circuit)

  • Small sample holding accessories

  • Reflection cupping plate (not removable)

  • Fiber test box

  • Film fixture

  • Micro transmission fixture

  • Rod box

Table Top Spectrophotometer

Labdex Table Top Spectrophotometer is designed for laboratory convenience with robust construction and high performance. It offers precise measurements, easy sample handling, and long-term stability. It is suitable for clinical laboratories, research facilities, and industrial testing.

Atomic Absorption Spectrophotometer

Labdex Atomic Absorption Spectrophotometer is designed for precise elemental analysis with high sensitivity and accuracy. It features stable light sources, advanced optical systems, and efficient background correction. It is widely used in environmental testing, pharmaceuticals, metallurgy, food safety, and chemical laboratories for trace metal analysis.

Double Beam Spectrophotometer

Labdex Double Beam Spectrophotometer ensures high photometric accuracy by continuously comparing sample and reference beams. It offers excellent wavelength stability, low noise, and automatic baseline correction. It is ideal for research laboratories, quality control, pharmaceutical, and biochemical analysis.

Touch Screen Double Beam Spectrophotometer

Labdex Touch Screen Double Beam Spectrophotometer combines advanced optics with an intuitive touch interface. It ensures high accuracy, real-time data display, and user-friendly operation. It is widely used in pharmaceutical analysis, biotechnology, and research laboratories.

Touch Screen Single Beam Spectrophotometer

Labdex Touch Screen Single Beam Spectrophotometer features a modern interface, fast operation, and reliable performance. It is ideal for routine absorbance measurements in educational laboratories, research centers, and quality control environments.